qLabs® FIB Test Strips
Sample type: Citrated whole blood
Sample volume: 15 μl
Test time: 20~600 seconds
Test range: 1~4g/L
Precision: CV ≤ 7%
INTENDED USE
The qLabs® test system is designed to provide quantitative measurement of fibrinogen (FIB) in sodium citrate whole blood sample. Fresh capillary whole blood, fresh venous whole and plasma shall not be used.
FEATURES
• Clot-based fibrinogen measurement compatible to Clauss fibrinogen method
• Automatic correction fibrinogen concentration by hematocrit
• Strong correlation to central laboratory results
• Dry-reagent strips stored at room temperature for immediate use
• Individually packed in humidity-resistant pouch
• External Quality Control
• Touchscreen handheld rapid testing platform with option of eStation II docking system
• Built-in QR code scanner for accuracy and efficiency
• Built-in communication port for printers and computer networks
PERFORMANCE
qLabs® FIB test shows very strong correlation with Stago STA®-Liquid Fib test. The correlation coefficient is 0.99
with an intercept of 0.0337 and slope of 1.0058 (n=98).
APPLICABLE DEVICES
qLabs Q-3 Plus
PACKAGE SPECIFICATIONS
24 tests/box
701-00022 qLabs.FIB.datasheet.A3.pdf