qLabs® PT/APTT Test Strips
Sample type: fingertip / fresh venous whole blood
Sample volume: 10 μl (one drop)
Test time: 3~7 minutes
Test range: INR: 0.5 - 8.0; aPTT: 20 - 420 sec
Precision: INR: ≤ 5.0 CV% aPTT: ≤7.0 CV%
INTENDED USE
The qLabs® Coag Panel 2 Test Strips are designed to provide quantitative determination for Prothrombin Time (PT)/International Normalized Ratio (INR) and Activated Partial Thromboplastin Time (APTT)
FEATURES
• Dry-reagent strips stored at room temperature for immediate use, no equilibration required
• Strong correlation to central laboratory results
• Smallest blood volume from fingerstick
• Stable external QC controls
• Built-in QR code scanner for accuracy and efficiency
• Built-in communication port for printers and computer networks
• Patented RACE TM technology for accurate aPTT determination
PERFORMANCE
qLabs® Coag Panel 2 test shows strong correlation with the Stago STA® NEOPLASTINE® CI PT-INR test. The correlation coefficient is 0.93 with an intercept of 0.11 and slope of 0.99.
qLabs® Coag Panel 2 test shows strong correlation with Stago STA® C.K. PREST® 5 aPTT test. The correlation coefficient is 0.95 with an intercept of -4.61 and slope of 0.966.
APPLICABLE DEVICES
qLabs® Q-3 Plus and qLabs® Q-3 Pro
PACKAGE SPECIFICATIONS
12 tests / box
701-00020 qLabs.Combo.datasheet.A4.pdf